Citadel Demonstration Video
This video is designed to provide users with an overview of the Citadel™ Patient Care System as well as some guidelines on its correct use.The new hospital bed Citadel™ Patient Care System provides flexible patient care for hospitals that demand safety, security and efficiency from their bed frames and therapeutic mattresses in Critical Care, High Dependency Units (HDUs) and Med Surg.
Integration provides important communication between the bed frame system and patient therapy system. One button delivers a CPR mode that simultaneously deflates the mattress and levels the frame, and automatic pressure compensation in the seat section occurs when the head of bed is adjusted. It also restricts automated turning when the head of bed exceeds 30 degrees or when one or more of the side rails is lowered.Support and Safety that only integration can provide
Powered therapeutic mattresses offer adaptive care for high risk patient profiles.- C100 Patient Therapy System
- C200 Patient Therapy System
Microclimate management
The Skin IQ® Family of compatible powered mattress covers utilize state-of-the-art Negative Airflow Technology (NAT) to continually draw away excess moisture from the skin/surface interface for microclimate management. Citadel offers a compatible Skin IQ outlet as standard, providing the user the option, and flexibility, to manage their patients microclimate effectively.Standardising efficiency
Citadel Patient Care System is part of the Arjo bed portfolio and is founded on a core design, built on a universal platform. Regardless of the bed frame chosen for your hospital, you have the assurance that each system is based on a consistent design, to help you provide optimal patient care.Optional Extras
- SafeSet Visual Alert (Brakes, Side rails, Low height, Bed exit armed)
- Foot Pedal (Deck Up/Down)
- Accessory Outlet
- Two way Nurse Call (US only)
- TV/ Light Controls (US only)
- RS232 Data Port for EMR (US only)
Standard
- Integrated Therapy Surfaces
- Tailored pressure ulcer prevention therapies to individual patient needs
- Flexibility in mattress choice (Active or Reactive therapies)
- Universal platform translating to lower lifetime costs
- IEC 60601-2-52 Safety Standards Certified by third party UL
- Practical infection control: easy cleaning and disinfection through openly accessible components (removable deck sheets)
- Light-weight and easy to maneuver
- Pit stop maintenance – self lubricating joints helps minimize preventative maintenance
- 20 degree lateral turn for patient repositioning
- Varizone patient movement detection & Bed Exit Alarm
- In Bed Weighing
- Under bed Anti Entrapment System (AES)
- One button CPR which delivers the CPR mode simultaneously to the mattress and the frame
- Ergonomic Attendant and Patient controls
- Function lockouts
- Automatic pressure compensation in the seat section when the head of bed is adjusted
- Pro-Contour Profiling
- Trendelenburg / Reverse Trendelenburg (min 12 degrees)
- Auto 30° backrest pause
- One button Cardiac Chair
- Underbed night light
- Skin IQ Outlet
- Linked brake castors
- 3 step bed extension
- Zero transfer gap (split sides)
- Battery back up
Technical Information
Safe working load | 270 kg (595 lb) |
Maximum patient weight | 227 kg (500 lb) |
Overall length Position 2 (standard) | 230 cm (90.6 in) |
Overall length Position 3 (extended) | 242 cm (95.3 in) |
Overall width | 103 cm (40.6 in) |
Mattress length Position 2 (standard) | 202 cm (80 in) |
Mattress length Position 3 (extended) | 214 cm (84 in) |
Mattress width | 88 cm (35 in) |
Mattress thickness | 15 cm to 20.5 cm (6 in to 8 in) |
Overall Product weight (approx.) | 200 kg (441 lb) |
Mattress and control unit weight (approx) | Mattress 16 kg (35 lb) Control unit 25 kg (55 lb) |
Height Range (with 125mm / 5in castors) | 34 cm to 78 cm (13.3 in to 30.7 in) |
Head / Foot down tilt angle | 12o min |
Backrest angle | 62o |
Backrest pause | 30o |
Thigh angle | 20o |
Calf angle | 16o |
Power input | 9.5A max at 115 VAC 60Hz 4.5A max at 230 VAC 50Hz |
Liquid ingress protection | IPX4 |
Electric shock protection | Class I type B |
EMC | Complies with EN 60601-1-2: 2002 and 2007 |
Safety standards |
ANSI/AAMI ES60601-1-2005 CAN/CSA C22 2 no. 60601-1-DB IEC 60601-2-52-2009 |